Olympus widefield and Confocal rescan


Purpose:
- Suitable for widefield fluorescence and transmission imaging
- Instant shift to confocal rescan imaging (slow). Pinhole size= 50 µm, Lateral resolution: 170 nm (FWHM at 488 nm excitation); Axial resolution: 500 nm (at 488 nm excitation); excitation: 405 - 488 - 561 - 638 nm
- Dedicated to short-term experiments, up to 4h

Illumination / Imaging configuration: Widefield / Confocal
Imaging contrasts: Brightfield, and Fluorescence
Microscope stand: Inverted Olympus IX81, motorized
Fluorescence excitation source WF: CoolLED pE-4000, LED light source 
Fluorescence excitation confocal: Toptica iChrome CLE lasers @20mW: 640nm; 561nm; 488nm; 405nm
Confocal band pass filters, MOUNTED ON Infinity optics filterwheel
(S1050/001/INF Cairn 1x C‐mount infinity port ):
ET450/50
ET525/36
ET600/50
ET655lp
Detection: 1 channel
Detection device: ORCA-Flash 4.0

Product number C133440-20CU
Imaging device sCMOS
Cell (pixel) Size (µm2) 6.5x6.5
Pixel array (horizontal by vertical) 2048x2048
Effective area (horizontal by vertical in mm) 13.312x13.312
Peak Quantum Efficiency (QE)  82% @560 nm
Dynamic range 37000 : 1


Acquisition software: INSCOPER: AUTOMATION & IMAGE ACQUISITION SOLUTION for Fluorescence Microscopy

https://www.inscoper.com/


Additional features:
XY motorized stage
- Hardware Z-drift compensation (Olympus ZDC2)

Objectives:

Objective type Mag/NA Medium Contrast WB (mm) Z-Drift compensation system Pixel size (nm)
UPlanFLN

20x / 0.5

Air

Fluo, Ph

2.1

/

397

PlanN

40x / 0.65

Air

Fluo

0.6

/

200

PlanApoN

60x / 1.42

Oil

Fluo

0.15

YES

133

UplanSApo

100x / 1.4

Oil

Fluo

0.13

YES

79


Fluorescence filter cubes installed:

Fluorophore

Excitation filter

Dichroic filter cut-off

Emission filter

Filter cube reference

DAPI

387/11

409

447/60

DAPI-1160B

GFP

482/18

495

520/28

GFP-1828A

YFP

500/24

520

542/27

YFP-2427B

mCherry

562/40

593

593LP

mCherry-40LP-A

Cy5

628/40

660

692/40

Cy5-4040B